British Journal of Oral and Maxillofacial Surgery, cilt.49, sa.2, ss.135-137, 2011 (SCI-Expanded)
Ten unembalmed adult sheep mandibles were used. The mandibles were sectioned in the midline, followed by sagittal split ramus osteotomies to obtain 20 hemimandibles. Each distal segment was advanced 5 mm on each hemimandible. Ten of the specimens were fixed with 4-hole extended 2.0 mm titanium miniplates and screws and the other 10 were fixed with 4-hole extended 2.0 mm locking miniplates/screws. Each fixed specimen was mounted on a servo-hydraulic testing unit with the fixation device, and was tested to a range of forces of 0-140 N. The displacement values (mm) under 20, 60, 120, and 140 N were compared with the help of the Mann-Whitney U-test, and there were no significant differences between them at any force tested. Locking miniplate/screws and standard miniplate/screws showed similar displacement values at the range of forces tested. © 2010 The British Association of Oral and Maxillofacial Surgeons.