Effect of radiation dose reduction on image quality in adult head CT with noise-suppressing reconstruction system with a 256 slice MDCT


Creative Commons License

Baskan O., EROL C., Ozbek H., Paksoy Y.

Journal of Applied Clinical Medical Physics, cilt.16, sa.3, ss.285-296, 2015 (SCI-Expanded) identifier identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 16 Sayı: 3
  • Basım Tarihi: 2015
  • Doi Numarası: 10.1120/jacmp.v16i3.5360
  • Dergi Adı: Journal of Applied Clinical Medical Physics
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.285-296
  • Anahtar Kelimeler: image quality, dose reduction, reconstruction methods, phantom studies, MDCT
  • İstanbul Medipol Üniversitesi Adresli: Evet

Özet

The purpose of our study was to investigate the effect of iterative reconstruction (IR) as a dose reduction system on the image quality (IQ) of the adult head computed tomography (CT) at various low-dose levels, and to identify ways of setting the amount of dose reduction. We performed two noncontrast low-dose (LD) adult head CT protocols modified by lowering the tube current with IR which were decided in the light of a group of phantom studies. Two groups of patients, each 100 underwent noncontrast head CT with LD-I and LD-II, respectively. These groups were compared with 100 consecutive standard dose (STD) adult head CT protocol in terms of quantitative and qualitative IQ. The signal-to-noise ratio (SNR) of the white matter (WM) and gray matter (GM) and contrast-to-noise ratio (CNR) values in the LD groups were higher than the STD group. The differences were statistically significant. When the STD and the LD groups were compared qualitatively, no significant differences were found in overall quality. By selecting the appropriate level of IR 34%, radiation dose reduction in adult head CT can be achieved without compromising IQ.